Error code
|
Area
|
text
|
Root cause
|
Action required
|
0
|
SGM II Gas analyser
|
Change O2 Cell
|
O2 cell expired
|
Replace O2 sensor and calibrate
|
1
|
SGM II Gas analyser
|
Pre_Warning O2 Cell near end of life!
|
O2 cell near end of life
|
Order replacement O2 cell ready to replace
|
2
|
SGM II Gas analyser
|
Temperature limit exceeded
|
Gas module overheated
|
check fan / filter on base of gas module
|
3
|
SGM II Gas analyser
|
BGL (Negative Drift)
|
Zero point drifted
|
ensure no fumes in workshop and zero calibrate unit
|
4
|
SGM II Gas analyser
|
Water Detected
|
water in gas sample system
|
allow SGM to automatically clear water
|
5
|
SGM II Gas analyser
|
Gas bench errors Present
|
Gas bench has fault
|
see other codes
|
6
|
SGM II Gas analyser
|
Auto zero calibration pending
|
Gas bench need to autocal
|
remove sample probe from exhaust and allow unit to autocalibrate
|
7
|
SGM II Gas analyser
|
Condensation in bench Detected
|
water in gas bench
|
leave pump running and allow sample system to dry out
|
8
|
SGM II Gas analyser
|
Low Gas Flow / Water in Pipe
|
filters blocked / water in pipe
|
allow water to clear / change filters/ blow hose out with an airline
|
9
|
SGM II Gas analyser
|
High Vacuum Reading
|
filters blocked / water in pipe
|
allow water to clear / change filters/ blow hose out with an airline
|
10
|
SGM II Gas analyser
|
Temperature Error
|
Gas bench overheating
|
check fan on gas module
|
11
|
SGM II Gas analyser
|
CO2 too high while HC residue check
|
Co2 high
|
remove probe from exhaust
|
12
|
SGM II Gas analyser
|
Gasbench driver error
|
USB connection problem
|
check USB connectors and or restart system
|
13
|
SGM II Gas analyser
|
No NO cell
|
no cell detected
|
connect cell
|
14
|
SGM II Gas analyser
|
No O2 cell
|
no cell detected
|
connect cell
|
15
|
SGM II Gas analyser
|
No Gas Flow / Pump not Running!
|
no flow detected from pump
|
check for blockage restart gas module
|
16
|
SGM II Gas analyser
|
No Acceptable new cell
|
new cell fitted outside of tolerance
|
check cell / replace as required
|
17
|
SGM II Gas analyser
|
Change NO cell
|
No cell needs changing
|
replace No cell
|
18
|
SGM II Gas analyser
|
NO cell requires calibration
|
No cell requires calibration
|
service engineer required
|
19
|
SGM II Gas analyser
|
Calibration Expired
|
Gas calibration expired
|
service engineer required
|
20
|
SGM II Gas analyser
|
Calibration due within 14 days
|
gas calibration due in 14 days
|
service engineer required within 14 days
|
21
|
SGM II Gas analyser
|
Leak Check Required
|
daily leak check required
|
perform leakcheck
|
|
|
|
|
|
245
|
SDP platform (PC)
|
Error cannot open the device OBD ROTI II
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
246
|
SDP platform (PC)
|
Error cannot open the device DSS 2 SMOKEMETER
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
247
|
SDP platform (PC)
|
Error cannot open the device REMOTE CONTROL
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
248
|
SDP platform (PC)
|
Error cannot open the device RPM & TEMPERATURE
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
249
|
SDP platform (PC)
|
Error cannot open the device GAS MODULE
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
250
|
SDP platform (PC)
|
Error cannot open the device DSS 10 SMOKEMETER
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
251
|
SDP platform (PC)
|
Error cannot open the device DSS 10 W SMOKEMETER
|
PC not seeing external device
|
check Bluetooth dongle location / configuration setting within SDP software
|
|
|
|
|
|
255
|
SDP platform (PC)
|
OBD ROTI II can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
256
|
SDP platform (PC)
|
DSS 2 SMOKEMETER can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
257
|
SDP platform (PC)
|
REMOTE CONTROL can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
258
|
SDP platform (PC)
|
RPM & TEMPERATURE can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
259
|
SDP platform (PC)
|
GAS MODULE can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
260
|
SDP platform (PC)
|
DSS 10 SMOKEMETER can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
261
|
SDP platform (PC)
|
DSS 10 W SMOKEMETER can't connect
|
PC not communicating with external device
|
check external device powered up / within operating range
|
|
|
|
|
|
330
|
DSS 10 W SMOKEMETER
|
Ambient temperature too high or too low.
|
Smokehead measuring air temperature out of range
|
Move head to a different location / engineer required
|
331
|
DSS 10 W SMOKEMETER
|
Optical detector temperature too high or too low.
|
Internal optic failure
|
service engineer required
|
332
|
DSS 10 W SMOKEMETER
|
The Tube Temperature is low.
|
internal measuring tube temp out of range
|
allow to warm up, ensure battery charge light is not on, service engineer required
|
333
|
DSS 10 W SMOKEMETER
|
Supply voltage too high or too low.
|
internal voltage out of range
|
service engineer required
|
334
|
DSS 10 W SMOKEMETER
|
Annual calibration required. Call for service !!
|
smokehead calibration required
|
service engineer required
|
335
|
DSS 10 W SMOKEMETER
|
Opacity out of range Zero calibration required.
|
Optic windows dirty or faulty
|
clean optics and try recalibrating unit
|
336
|
DSS 10 W SMOKEMETER
|
Requested number of sample points have not been sampled completely.
|
insufficient smoke to complete test
|
retest vehicle
|
337
|
DSS 10 W SMOKEMETER
|
Smoke head has been restarted. Zero calibration required !
|
unit has restarted and not done self test
|
perform a zero calibration
|